Clarius Scanner HD3 Disinfecting
revision 9 39
To disinfect your scanner (high-level):
1. Make sure the fan is still detached from the scanner.
2. Mix the disinfectant solution by following the disinfectant label instructions for
solution strength and disinfectant contact duration.
3. Using a compatible disinfectant at a temperature of 23°C (73°F), immerse the scanner
in the disinfectant solution as per the recommendation provided within the
disinfectant instructions.
4. Using the instructions on the disinfectant label, rinse the scanner.
5. Air-dry Alternatively, use a non-linting cloth.
6. Examine the components for damage, such as cracks or splitting where fluid can enter.
If damage is evident, do not use the scanner and contact Clarius Mobile Health.
Disinfecting the Clarius Power Fan HD3
Before you begin disinfecting, make sure you have cleaned the fan (see Cleaning the Clarius
Power Fan HD3 on page 37).
Because the Clarius Power Fan HD3 cannot be submerged in liquid, never use high-level
disinfectants. Always use intermediate-level disinfectants. For a list of disinfectants
recommended for intermediate disinfection of the Clarius Power Fan HD3, refer to Cleaners &
Disinfectants on page 74.
To disinfect the Clarius Power Fan HD3:
1. Make sure that the fan is detached from the scanner.
2. Disinfect the fan by wiping with a cloth moistened with a compatible disinfectant.
Alternatively, use a premoistened disinfectant wipe.
3. Air-dry. Alternatively, use a non-linting cloth.
4. Examine the fan for damage, such as cracks or splitting. If damage is evident, do not
use the fan and contact Clarius Mobile Health.
Disinfecting the Micro USB Cable
To disinfect the Micro USB cable:
1. Make sure the Micro USB cable is detached from the power supply and the Clarius
Power Fan HD3.
2. Disinfect the Micro USB cable by wiping with a cloth moistened with a compatible
disinfectant. Alternatively, use a premoistened disinfectant wipe.
3. Air-dry. Alternatively, use a non-linting cloth.