Clarius Scanner HD3 Imaging
revision 9 25
1. After three seconds, it decreases frame rate.
2. After 30 seconds of decreased frame rate, it freezes.
3. After 10 seconds in freeze mode, it goes idle.
4. After 15 minutes of idle time, it shuts down.
When you go to the image acquisition page to begin an exam, the Clarius Scanner HD3
automatically switches from standby mode to scanning mode. For instructions on using the
imaging tools, go to support.clarius.com/hc/en-us.
A typical use of the Clarius Scanner HD3 is described as five continuous scanning minutes
followed by 10 minutes in standby mode (or turned off).
• Notifications and alerts from third-party applications may interrupt you or the Clarius App,
thereby interfering with the exam. Configure your smart device in accordance with your
institution's security policies.
• Vibration range that is too high for the scanner may cause the scanner to malfunction
during an exam.
• Using improper gel type or combining different gel types may expose patients to risks and
produce poor-quality images.
• When using PAL HD3 with dual array technology, there will be an offset on the transducer
physical placement in the elevation axis. Due to the side-by-side physical arrangement of
the transducers, refer to the transducer elevation width center line for the linear array
when doing ultrasound guided injection procedure.
• When using PAL HD3 with dual array technology, the preset selection will determine array
utilization. Care must be taken to avoid using the wrong preset when doing Ophthalmic
imaging (low MI/TI limit). Ensure appropriate preset has been selected before imaging
begins.
For proper transmission of the acoustic beam, use only Aquasonic 100, and use it only before
its expiry date. Download the usage instructions from parkerlabs.com and read all the
information before operating the device.
Do not use:
• Lotion-based products or gels that contain mineral oil.
• Hand-sanitizing gels.
• Scanners left soaking in gel.
When scanning for more than 15 consecutive minutes, with any Clarius Scanner HD3, the fan
must be used. For information on the fan, see Clarius Power Fan HD3 on page 32.