Clarius Scanner HD3 Safety Topics
revision 9 61
• In Canada, the device for operation in the band 5150 – 5250 MHz is only for indoor use to
reduce the potential for harmful interference to co-channel mobile satellite systems.
Note: Settings for channel band selection of Clarius Ultrasound Scanner is managed by Clarius
Mobile Health Corp. as per local wireless regulation. Channel band being selected can be
found under the App "Status" page. Users need to ensure channel band usage is in compliance
with indoor and outdoor band allowance as per local wireless regulation.
Electrostatic Discharge Precautions
Electrostatic discharge (ESD), or static shock, results from the flow of an electrical charge from
a person or object of a higher charge to that of a lower charge. ESD is most prevalent in low-
humidity environments, often caused by heating or air-conditioning.
To reduce ESD:
• Use anti-static spray on carpets, linoleum, and mats. Or use a ground wire connection
between the system and the patient table or bed.
Electromagnetic Emissions
Ensure that the Clarius Ultrasound Scanner is used only in those operating environments
indicated in the following table. Operating the system in an environment that does not meet
these conditions may degrade system performance.
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by
DECLARATION OF ELECTROMAGNETIC EMISSIONS
Emissions Test Compliance Electromagnetic Environment
RF emissions, CISPR 11 Group 1 The system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions, CISPR 11 Class B The system is suitable for use in all establishments,
except domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions, IEC 61000-3-2 Class A
Voltage fluctuations/flicker emissions, IEC
61000-3-3
Complies