Software ver. CVPCA30E_1 29
Repeat the last step as required in order to edit the protocol name. To
insert a space in text position the cursor in the top left blank space of the
grid and press . The maximum name size is 40 letters (20 on the
upper row and 20 on the lower row) including spaces.
When you have finished building your protocol name, position the cursor in
the top left blank space and press
To enter values for this protocol ensure it is highlighted and press .
Enter VTBI in ml. to change any numeric program parameter just enter the
numbers which will replace the previous values and this will overwrite the
previous value. If you make a mistake press which will delete the last
digit. Press when correct value is entered.
Change the initial rate using the numeric
keypad. If the rate step does not appear then
the maximal rate under "Change Set Up" has
been set to 0ml/hr to prevent users from
delivering regimes with a continuous element.
If you have the authority to change this go to
"Change Set Up" (see later section). If the
desired rate for the protocol you are editing is
in ml/hr then enter the rate and press to
accept. Press to confirm.
NOTE: The BodyGuard® 575 is capable of continuous (basal) rates of up
to 30ml/hr. User can enter a maximum basal rate in "Change Set Up" using
the level three access-code to ensure other users can not accidentally
program the pump to deliver above the safe ceiling for your commonly
used drugs.
To specify the concentration in mg or mcg/ml
of the drug regime you are programming,
press 3 times in the rate menu until you
see the concentration screen.