Coil has disconnected from the implant
The coil is no longer picking up a signal from the
implant. This alert could mean that:
The • coil has fallen off your head.
There may be a • fault with the coil or coil
cable.
Suggested actions:
Step 1: Place the coil back on your head.
Make sure the coil magnet is properly positioned over the
implant.
If the coil falls off your head often, try:
• Adjusting the magnet depth, or
Changing to a stronger magnet.•
If this solved the problem, you’re done. If not, continue.
Step 2: Check both the coil and the coil cable as described in
Perform a maintenance check.
If this solved the problem, you’re done. If not, continue.
Step 3: Contact your clinician.
14 - CP810 and CR110 Troubleshooting Guide © Cochlear Limited 2010
Use your remote assistant to troubleshoot your processor