21. Specifications
Mechanical
• Dimensions: L:140 mm x B:140 mm x H:62 mm.
• Weight: 499 g
Operational
• Flow Rate: 1 mL/h to 600 mL/h
• Volume: 1 mL to 4000 mL
• Increments: 1 mL
• Feed Until Empty (Continuous) CONT
• 3ULPLQJ5DWH P/K
• Accuracy: ± 5% according to the following parameters:
Rate: 100 mL/h
Volume: 100 mL
Liquid: Water
Administration Set: New, previously unused set
Up to 1500 mL delivered volume*
*Delivery of more than 1500 mL with one set may affect accuracy.
Head Height above rotor: 30cm
• +HDG+HLJKW(IIHFWVRQ$FFXUDF\ SHUFPRIKHDGKHLJKW
Head height range: -30cm to 60cm
• %DFN3UHVVXUH(IIHFWV SHUN3D
• Occlusion Detection Pressure:
Upstream: -34 kPa for 5 seconds
Downstream: 83 kPa for 30 seconds
103 kPa for 5 seconds
Note: Low flow rates may significantly extend the time to occlusion alarm
• Air Detection 1.5 mL
• Operating Temperature: +5° C to +40° C
• Operating Humidity: 15% to 93%, non-condensing
• 2SHUDWLQJ$WPRVSKHULF3UHVVXUH WRN3D
Storage
• Storage Temperature: -20°C to +60°C
• Storage Humidity: 10% to 95%, non-condensing
• Atmospheric Pressure: 50 to 106 kPa
Power
• Charger:
Input: 100-240 V; 50-60 Hz
Output: 5 VDC 2A, 10.0 W
Pump:
Input : 5 VDC 2A, 10.0 W
• Fuse Type: Single blow fast acting
• Battery Type: Lithium Ion
• %DWWHU\&KDUJH7LPH KRXUV
• Battery Life: 24 hours, at 125 mL/h
• Battery capacity decay due to:
Storage: For optimal battery life, do not expose battery to elevated
storage temperatures for extended periods. Store below
50°C when possible and do not exceed 60°C storage
temperature. For extended storage, store battery at low
temperature and low charge state.
Aging: Please note that the battery capacity numbers noted below
are estimates and that actual battery capacity may vary sig-
nificantly depending on factors such as actual use, charge
cycles and temperature.
EN
ƐƐ
•
C
M
Y
CM
MY
CY
CMY
K