EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
Annual System Maintenance
Preventive maintenance, safety, power, and calibration checks should be performed annually by a
Cutera representative to ensure proper laser performance.
System Repair
All system repairs should be performed by a Cutera-certified service engineer. For training and
information, contact your local Cutera representative.
Routine User Maintenance
Cleaning the External Surfaces of the System Console
Use a cloth dampened with non-caustic cleaning solution, such as mild soap and water, isopropyl alcohol
or a “hospital-grade” disinfectant, to wipe the external surfaces of the system console. Dry with a clean
cloth or allow to air dry.
Cleaning the System Touchscreen Display
Apply an alcohol-based cleaner to a soft cloth to clean the system touchscreen display.
Clean the external surfaces of the system console Weekly, or as needed
Clean the system touchscreen display Weekly, or as needed
Inspect and clean the handpiece During treatment, as needed, and between
patient use
Inspect and clean the heat exchanger intake Inspect weekly; clean monthly, or as needed
Fill the coolant reservoir On installation or if water evaporates
WARNING
Do not attempt to gain access to any internal components. Electrical
shock and/or unintended laser exposure may result. Do not spray or
pour cleaning agents directly on the system console.
CAUTION
Do not spray or pour cleaning agents directly on the system console
or control screen. You may damage the console, touchscreen and
the system electronics.