69
EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
Physical Parameters
Size (W X D X H): 16.1 in X 24 in X 34.8 in (.41 m x .61 m x .88 m)
Weight: 185 lbs (84.1 kg)
Operating length of umbilical: 9 ft (2.7 m)
Minimum bend radius of umbilical: Storage 5.2 in (132 mm)
Momentary 2.6 in (66 mm)
Electrical Requirements
Voltage, Current 100-120 VAC, 20 A or 200-240 VAC, 16 A
Frequency 50/60 Hz
Environmental Requirements
Temperature Operating 15º C to 30º C
Non-operating -10º C to 50º C, coolant drained
Humidity 0 to 90%, non-condensing
Max. Heat Dissipation 4,000 Watts
Max. Altitude Operating 15,000 ft
Non-operating 50,000 ft
Atmospheric Pressure Operating: 60 to 110 kPa
Maximum Allowable Temperature of Accessible Parts
Parts User/Patient Can Touch Material Duration Limits
Enclosure Plastic > 1 min 48º C
Touchscreen Plastic > 1 min 48º C
Coolview Handpiece Window Sapphire > 1 min 20º C
CoolView Handpiece Treatment Guide Metal > 1 min 20º C

Table of Contents