EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
Clean the handpiece lens(es) using a clean wood stick cotton swab and lens cleaning grade acetone, as
described below.
1. Ensure that the system is turned OFF or is in STANDBY mode.
CAUTION
Inspecting and cleaning the handpiece lens(es) is the responsibility
of the practitioner. If the lens(es) is (are) not properly maintained,
debris can build up on the surface of the lens(es), leading to
permanent damage to the lens(es) and, ultimately, permanent
damage to the handpiece optical system. If debris cannot be
removed from the lens(es), the handpiece must be replaced.
CAUTION
Do not use germicidal disposable wipes to clean the recessed
lens(es). Doing so may damage the lens(es) and require handpiece
replacement.
NOTE
Debris from patients can accumulate on the lens(es). It may be
necessary to clean the lens(es) periodically during treatment.
NOTE
It may be necessary to use a flashlight to inspect the Genesis V
handpiece lenses.
NOTE
Remove the Dermastat handpiece cap to facilitate inspection and
cleaning of the handpiece lens. Be sure to replace the cap before
use.
NOTE
For optimal results, use acetone with 0.5% or lower water content,
which is available at most camera shops. If unable to obtain acetone
with 0.5% or lower water content, contact Cutera Service (refer to
Appendix D).
NOTE
Use only wood stick cotton swabs. Do not use products such as Q-
Tips as they contain an adhesive that dissolves when it comes in
contact with acetone.