88
EXCEL V+ OPERATOR MANUAL
D2325, REV A, 2/19
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The excel V+ is intended for use in the electromagnetic environment specified below. The customer or the user
of the excel V+ should ensure that it is used in such environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment:
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8kV contact
±15kV air
±2, 4, 6, 8kV contact
±2, 4, 8, 15kV air
Floor should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30-60%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kV for power sup-
ply lines
±1kV for input/output
lines
±2kV line to ground
±1kV line to line
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV differential
mode
±2kV common mode
±0.5, 1kV
differential mode
±0.5, 1, 2kV
common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
0% U
T
.5 cycle
0% U
T
1 cycle
70% U
T
25 cycles
0% U
T
5 Sec
0% U
T
.5 cycle
0% U
T
1 cycle
70% U
T
25 cycles
0% U
T
5 Sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the excel V+
requires continued operation during
power mains interruptions, it is
recommended that the excel V+ be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power-frequency magnetic fields
should be at levels characteristics of a
typical location in a typical commercial
or hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.