23
8-5. SPAN constant claibration
It is to calibrate with the S.CAL value written down for load calibration. You can
calibrate without any standard weight.
Measuring mode
While key is being
pressed, Press
key
Displays Function Mode State
Displays Sensor Ouput Value Calibration
Displays Actual Load Calibration State
Input the rated output value by using and key. And
then, press
key. (In
p
ut 3.0000 if donot know the rated
output of sensor)
Calibration Completed
Displays Shunt cal value Calibration state
Displays the Rated output value input state
Displays Shunt cal value input State
Input S.CAL value by using
and key. And then
p
ress
key