AquaLab
2. About AquaLab
8
Chilled Mirror Dewpoint Limitations
AquaLabs limitation is its ability to accurately measure samples with
high concentrations (typically >1%) of certain volatiles such as etha-
nol or propylene glycol, which can condense on the surface of the
chilled mirror.  e extent of the e ect is determined by how readily
the material volatilizes, which is both concentration- and matrix-
dependent.  erefore, even if your sample contains materials that
could volatilize, it may still be possible to make accurate readings
using the chilled mirror dewpoint sensor.
AquaLab Series 4TEV which incorporates both a chilled mirror sen-
sor and a capacitance sensor for measuring volatile substances is
Decagons solution for products containing volatile materials. If you
are unsure if you need the TEV model, please call and discuss your
product with a Decagon Representative. Refer to Chapter 8’s section
titled ”Volatile Samples” or call Decagon for more details.