AquaLab
9. Taking a Reading
52
9. Taking a Reading
Measurement Steps
Once you have veri ed for cleanliness, calibration and prepared
your sample, you are ready to take readings.  e process is simple:
Move the chamber lever to the Open position and lift the cham-
ber lid.
Check the top lip and outside of the sample cup to make sure
they are free from sample residue and that sample cup isnt over-
filled. (remember, an over-filled sample cup may contaminate
the chambers sensors).
Place your prepared sample cup in the chamber. The sample cup
lid must be removed while in the testing chamber for correct
functionality.
Close the chamber lid and move the lever to the Read position.
This will seal the chamber and start the reading.
In 1 to 2 minutes, the  rst water activity measurement will be dis-
played on the LCD (this is an intermediate reading and not the
nal water activity). Length of read times may vary depending on
temperature di erences between the chamber and your sample, and
other properties of your sample.
How AquaLab Takes Readings
AquaLabs reading cycle continues until the rate of change of three
consecutive readings are less than 0.0005 a
w
of each other.  e in-
strument crosses the dew threshold numerous times to ensure equi-