AquaLab
6. Cleaning and Maintenance
33
Purpose
e purpose for the cleaning procedure is to remove grease, dirt and
other soluble substances which can absorb/release water during veri-
cation, calibration, and/or sample testing. For a smooth and even
dew formation, it requires the mirror to be perfectly clean. If there
are any contaminants (e.g. ngerprints) on the mirror, the dew will
form unevenly and thus a ect the accuracy of the reading.
When to Clean
e instrument should be cleaned if visual inspection indicates the
chamber is dirty or as instructed in the veri cation owchart on
page 41.
Materials Needed
• A thin plastic rod or other non-metal implement
• Distilled Water
• Isopropyl Alcohol (IPA) or Decagon Cleaning Solution
• Kimwipes®
You may also purchase the AquaLab Cleaning Kit which comes with
all the above materials except the Isopropyl Alcohol and Distilled
Water.
NOTE: Wash your hands with soap and water and/or use clean lab gloves
before starting the cleaning procedure. is will prevent oils from contam-
inating the cleaning materials, the sample chamber and/or the sensors.
Cleaning the Block and Sensors
Accessing the Sample Chamber
Turn the power o on your AquaLab. If latched, move the lever
over to the open position. Lift the chamber cover to expose the