AquaLab
12. Troubleshooting
76
7. PROBLEM
Veri cation is not correct.
SOLUTIONS:
1)  e sample chamber and mirror need to be cleaned. See
Chapter 6 for detailed cleaning instructions. If veri cation is
still not correct, then linear o set has occurred.
2) Verify and Adjust for Linear O set. After you have cleaned
the sample chamber and mirror (Chpt. 7) you will need to
use a Veri cation Standard to verify and adjust for Linear
O set as described in Chapter 7.
8. PROBLEM:
Message on screen displays the following:
SOLUTION:
e crystal that runs the  rmware is having trouble starting. Oc-
casionally, cycling the power will solve the problem. If this mes-
sage continues to appear, the instrument will need to be ser-
viced by Decagon. See Chapter 13 for detailed instructions.