14.3.11 CAlibration:VOLTage:MEASure:OFFSet
Used to calibrate the zero-offset of the voltage monitoring.
CAlibration:VOltage:MEASure:OFfset?
Reads-back the present offset calibration value of the voltage monitoring.
14.3.12 CAlibrate:VOltage:MEasure:GAin
Used to calibrate the full-scale of the voltage monitoring.
CAlibrate:VOltage:MEasure:GAin?
Reads-back the present full scale calibration value of the voltage monitoring.
14.3.13 PAssword
From factory, the password is set to “DEFAULT” and *SAV can be used without parameter.
With the command PAssword<wsp>DEFAULT,<New_password> the protection is activated.
“DEFAULT” is the factory default string, <New_password> is the new password.
The string is maximum 8 characters and is not case sensitive. Execution of the *SAV command without
valid password causes an execution error 15 (Illegal password).
A few examples:
PA DEFAULT,MYPASSW<LF>
Changes the factory default password into a new password.
After this command the *SAV command works only with a valid password as parameter.
The command:*SAV MYPASSW<LF>.
PA <current_password>,<new_password>
Changes the current password into a new password.
PA MYPASS,DEFAULT
Changes the current password into the factory default. Protection is disabled.
PAssword?
This query returns a 0 if the access is not protected. Otherwise a 1 is returned when the acces is
protected with a password.
Forgotten password. With the command PAssword:Reset or abbriviated PA:R the PSC comming into a
non calibrated state and the password are set to “DEFAULT”.
PSC SERIES DELTA ELEKTRONIKA BV
May 2008 MANUAL page 26