14.3 SCPI commands
14.3.1 SOurce:VOltage:MAximum
14.3.2 SOurce:CUrrent:MAximum
These commands are used to enter the maximum voltage and maximum current range of the used
power supply. For example, the Delta power supply model SM35-45 has a range of 35 volt and 45
ampere. The SOur:VOlt:MAx<wsp>35 and the SOur:CUrr:MAx<wsp>45 have to be given before the
voltage and current set commands. Save the configuration values in non-volatile memory with the *SAV
command. The next time settings will automatically recalled at power on.
SOur:VOlt:MAx? and SOur:CUrr:MAx? read back the voltage and current settings.
14.3.3 CAlibration:CUrrent:OFfset
CA:CU:OF<wsp><NR1>
Used to calibrate the zero-offset of the current programming.
14.3.4 CAlibration:CUrrent:OFfset?
Reads-back the present offset calibration value of the current programming.
14.3.5 CAlibration:CUrrent:GAin
Used to calibrate the full-scale of the current programming.
14.3.6 CAlibration:CUrrent:GAin?
Reads-back the present full scale calibration value of the current programming.
14.3.7 CAlibration:VOltage:OFfset
Used to calibrate the zero-offset of the voltage programming.
CAlibration:VOltage:OFfset?
Reads-back the present offset calibration value of the voltage programming.
14.3.8 CAlibration:VOltage:GAin
Used to calibrate the full-scale of the voltage programming.
CAlibration:VOltage:GAin?
Reads-back the present full-scale calibration value of the voltage programming.
14.3.9 CAlibration:CUrrent:MEasure:OFfset
Used to calibrate the zero-offset of the current monitoring.
CAlibration:CUrrent:MEasure:OFfset?
Reads-back the present offset calibration value of the current monitoring.
14.3.10 CAlibration:CUrrent:MEasure:GAin
Used to calibrate the full-scale of the current monitoring.
CAlibration:CUrrent:MEasure:GAin?
Reads-back the present full scale calibration value of the current monitoring.
page 25 MANUAL May 2008
DELTA ELEKTRONIKA BV PSC SERIES