14.3.29 CU
With the CU command you can change the custom string of the identity string. See chapter 14.2.20.
The string must have 15 characters including the <lf> and is send via RS232.
Some examples:
CU<WSP>CAL. 01-02-2003 Calibration date
CU<WSP>SM 70-20 no 007 Power supply type and inventory number
After this command you can check it with query *IDN? to display the identify string.
With *SAV you can store the new information in non volatile memory.
PSC SERIES DELTA ELEKTRONIKA BV
May 2008 MANUAL page 30