EasyManua.ls Logo

Despatch LCC Series - Figure 22. Adjust Nitrogen Needle Valve (in Compartment Beneath Oven); Table 7. Desired Oxygen Concentration

Default Icon
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LCC_LCD1-16/51 Oven Owner’s Manual C244, 320207, REV B
59 of 82
b. Press or to display the desired profile.
c. Press Run to start the Profile Mode.
6. The first segment of the program is the PURGE MODE.
a. Set the nitrogen flow meter (Figure 21) to the desired setting (Table 7).
b. This program energizes the purge solenoid valve.
Table 7. Desired Oxygen Concentration.
Oven Model
Mode
< 100 PPM
<1000 PPM
LCC1-16
Purge
150 SCFH
125 SCFH
Maintain
75 SCFH
65 SCFH
LCC1-51
Purge
200 SCFH
175 SCFH
Maintain
160 SCFH
140 SCFH
7. The second segment is the MAINTAIN MODE:
a. The purge solenoid valve is de-energized and the maintain solenoid is
energized to maintain the nitrogen level to less than the purge level.
b. Adjust the Nitrogen Maintain needle valve at rear of oven (Figure 22).
Leave the maintain valve energized as long as the nitrogen level is
maintained (Table 7).
8. The third and final segment is the COOLDOWN MODE. In Cooldown mode,
water valves are energized to bring the chamber to a safe unloading
temperature (Figure 23).
a. Set water cooling flow meter to 2 GPM (11 LPM) and adjust if necessary.
Figure 22. Adjust Nitrogen needle valve (in compartment beneath oven).

Table of Contents

Related product manuals