TP-700 Instruction Manual Rev. 4.1 Page 30 of 42
Caution: A hazardous area must be declassified before opening the junction box or
removing and replacing the plug in sensor
a) Open the junction box and remove power to TP-700 sensor by lifting the + 24VDC wire in J-Box.
b) Use a M1.5 Allen wrench to release the locking setscrew that locks the ITM and bottom housing together
(One turn will suffice - Do not remove setscrew completely).
c) Remove splashguard. Unthread and remove the Bottom Housing from the ITM.
d) Gently pull the plug-in H
2
S sensor cell out of the ITM. Orient the new plug in sensor so that it matches with
the female connector pins. It may be necessary to look from below to assure alignment is correct. When properly
aligned, press the sensor in firmly to make the proper connection.
NOTE: The previous plug-in H
2
S sensor cell did not use a Face seal o-ring design. If you are
installing this new face seal o-ring version to replace an older revision sensor, you must fully
remove the adhesive gasket from the inside of bottom housing first
e) Thread the Bottom Housing onto the ITM to a snug fit and tighten the locking setscrew using the 1/16” Allen
wrench. Reinstall the splashguard.
f) Perform “Set Heater Power (Section 3.5.6 to match the new sensor with the ITM.
g) Perform a successful AutoSpan to match the new sensor with the ITM (Section 3.4).
5.5 Replacement of ITM
Caution: Hazardous areas must be declassified before opening the junction box or
removing and replacing the ITM
a) Remove the power source from the sensor assembly. Disconnect all sensor wire connections at the J-Box
taking note of the wire connections.
b) Use a wrench and loosen the locking nut at the top of the ITM and unthread the ITM from the junction box.
c) Use a M1.5 Allen wrench to release the locking setscrew that locks the ITM and bottom housing together
(One turn will suffice - Do not remove setscrew completely).
d) Remove splashguard. Unthread and remove the Bottom Housing from the ITM.
e) Gently remove the plug-in H
2
S sensor from the old ITM and install it in the new ITM. Orient the plug-in
sensor so that it matches with the female connector pins on the new ITM and press the sensor in firmly to make
proper connection.
f) Thread the bottom housing onto the ITM until snug, tighten the locking setscrew and reconnect splashguard.