EasyManua.ls Logo

Detecto 750

Detecto 750
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SETUP AND CALIBRATION, CONT.
Unit=
(WEIGHING UNIT)
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI/ENTER key to save it and proceed to the next
prompt. Allowable values are: 1, 2, 3 or 4.
1 = Pounds Only 2 = Kilograms Only
3 = Pounds/Kilograms 4 = Kilograms/Pounds
5 = Pounds & Ounces/Kilograms (if USA=1)
UnS=
(MOTION (UNSTABLE) RANGE)
The Motion (Unstable) Range is the number of divisions of change permitted before the
STABLE annunciator turns off.
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys and the _ or ` keys to
scroll through and select a new value and then press the BMI/ENTER key to save it and
proceed to the next prompt. Allowable values are 0 through 99.
dPP=
(DECIMAL POINT POSITION)
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI/ENTER key to save it and proceed to the next
prompt. Allowable values are: 0, 1, 2, or 3.
0 = XXXXXX 1= XXXXX.X 2 = XXXX.XX 3 = XXX.XXX
CAP=
(SCALE CAPACITY)
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys and the _ or ` keys to
scroll through and select a new value and then press the BMI/ENTER key to save it and
proceed to the next prompt. Allowable values are: 1 through 99,999.
FLT=
(DIGITAL FILTER LEVEL SELECTION)
Your indicator will arrive with factory filter settings of 1 = Minimal. Please check with Tech
Support before changing filter level, break range and sample rate.
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI/ENTER key to save it and proceed to the next
prompt. Allowable values are:
0 = Minimal Filter 1 = Moderate Filter 2 = Heavy Filter 3 = Custom Filter
NOTE! If 3 = Custom Filter is selected, two additional prompts will be displayed.
8555-M483-O1 Rev A x 750 Owner’s
14

Other manuals for Detecto 750

Related product manuals