SETUP AND CALIBRATION, CONT.
Unit=
(WEIGHING UNIT)
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI/ENTER key to save it and proceed to the next
prompt. Allowable values are: 1, 2, 3 or 4.
1 = Pounds Only 2 = Kilograms Only
3 = Pounds/Kilograms 4 = Kilograms/Pounds
5 = Pounds & Ounces/Kilograms (if USA=1)
UnS=
(MOTION (UNSTABLE) RANGE)
The Motion (Unstable) Range is the number of divisions of change permitted before the
STABLE annunciator turns off.
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys and the _ or ` keys to
scroll through and select a new value and then press the BMI/ENTER key to save it and
proceed to the next prompt. Allowable values are 0 through 99.
dPP=
(DECIMAL POINT POSITION)
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI/ENTER key to save it and proceed to the next
prompt. Allowable values are: 0, 1, 2, or 3.
0 = XXXXXX 1= XXXXX.X 2 = XXXX.XX 3 = XXX.XXX
CAP=
(SCALE CAPACITY)
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys and the _ or ` keys to
scroll through and select a new value and then press the BMI/ENTER key to save it and
proceed to the next prompt. Allowable values are: 1 through 99,999.
FLT=
(DIGITAL FILTER LEVEL SELECTION)
Your indicator will arrive with factory filter settings of 1 = Minimal. Please check with Tech
Support before changing filter level, break range and sample rate.
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI/ENTER key to save it and proceed to the next
prompt. Allowable values are:
0 = Minimal Filter 1 = Moderate Filter 2 = Heavy Filter 3 = Custom Filter
NOTE! If 3 = Custom Filter is selected, two additional prompts will be displayed.
8555-M483-O1 Rev A x 750 Owner’s
14

Other manuals for Detecto 750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Detecto 750 and is the answer not in the manual?

Detecto 750 Specifications

General IconGeneral
Readability0.2 lbs
DisplayLCD
PowerAC adapter or battery
ConnectivityRS232
FunctionsBMI

Related product manuals