SETUP REVIEW
The 750 indicator allows several operational parameters to be reviewed and changed as
necessary without having to enter the setup mode.
To Enter Setup Review:
1. If indicator is on, press ON/OFF key.
2. Display will show
and indicator will turn off.
3. Press and hold BMI/ENTER key and then press ON/OFF key.
4. Indicator will perform a lamp test (turn on all segments of display), display model
number and software revision and then display
prompt.
5. With display showing
, release BMI/ENTER key.
6. Refer to instructions listed in Setup and Calibration section for information on how to
change parameters.
Parameters in Setup Review will be processed in the following sequence:
Select the number of scale divisions automatically zeroed on power up
Enable or Disable automatic reset of weight display to zero on power up
Select baud rate for serial output
Select to use Ni-Cad, NiMH or Alkaline batteries and type of charging
Disable or select number of minutes of inactivity at zero for sleep mode
Disable or select number of minutes for automatic shutoff timer
EVENT COUNTER
The Model 750 Weight Indicator has been designed with an Event Counter type of security
seal. When selected, the 750 will display two 3-digit numbers representing the Calibration and
Configuration counters.
Calibration Counter
The calibration counter is incremented when a value in the calibration part of setup is
changed (
,
,
,
,
,
,
,
,
,
,
,
,
,
).
The counter is only incremented 1 time even if more than one parameter is changed each
time through setup.
Configuration Counter
The configuration counter is incremented when a value in the configuration part of setup is
changed (
,
,
,
,
,
,
,
). The counter is only
incremented 1 time even if more than one parameter is changed each time through setup.
To Review the Event Counter:
1. If indicator is on, press ON/OFF key.
2. Display will show
and indicator will turn off.
3. Press and hold the UNITS key and then press ON/OFF key.
4. The display will then show
(Calibration Check) for two (2) seconds followed by
two 3-digit numbers.
x The left number represents the setup Configuration counter.
x The right number represents the Calibration counter.
5. To return to the normal operation, press the UNITS key.
6. Otherwise press the ON/OFF key to turn off the indicator.
8555-M483-O1 Rev A x 750 Owner’s
19