SETUP AND CALIBRATION, CONT.
F=
(FILTER LEVEL)
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI/ENTER key to save it and proceed to the next
prompt. Allowable values are 1
(least amount of filtering) to 99 (greatest amount of filtering).
BR=
(BREAK RANGE)
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI/ENTER key to save it and proceed to the next
prompt. Allowable values are 1
to 99 which correspond to the number of division changes
to break out of filtering.
SR=
(SAMPLE RATE)
Press the BMI/ENTER key to view the current setting. If the value displayed is acceptable,
press the BMI/ENTER key again. Otherwise, press the a or b keys to scroll through and
select a new value and then press the BMI/ENTER key to save it and proceed to the next
prompt. Allowable values are a minimum of 1
sample per second to a maximum of 16
samples per second in one sample per second intervals.
CAL=
(CALIBRATION)
With the display showing
CAL=
, press the BMI/ENTER key. The display will change to show
the current setting
0
(0=NO). If the scale has been previously calibrated and you wish to skip
calibration and proceed to the
HOLD=
(Hold Mode) prompt, press the BMI/ENTER key and
the previous calibration will be retained.
To begin calibration, press the a key to select 1 (yes) and then press the BMI/ENTER key.
After pressing the BMI/ENTER key, the display will change to
LOAd=
.
LoAd=
(LOAD CALIBRATION WEIGHT)
With the display showing
LoAd=
perform the following steps:
1. Make certain the scale platform is empty and free of debris.
2. Place the desired amount of calibrated test weights on the scale platform. A minimum
of 50% of scale's capacity is required. However 70% to 100% is recommended.
3. Press the BMI/ENTER key.
4. If the value displayed is acceptable, press the BMI/ENTER key again. Otherwise,
determine the exact amount of test weight placed on the scale platform and then using
the a or b keys and the _ or ` keys scroll through and select the test weight amount.
5. Verify that the numbers selected are the same as the amount of the test weight and
then press the BMI/ENTER key.
6. Starting at the left and preceding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the display
will proceed to the next prompt.
8555-M483-O1 Rev A x 750 Owner’s
15

Other manuals for Detecto 750

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Detecto 750 and is the answer not in the manual?

Detecto 750 Specifications

General IconGeneral
Readability0.2 lbs
DisplayLCD
PowerAC adapter or battery
ConnectivityRS232
FunctionsBMI

Related product manuals