3300-0188-0M Rev L y apex Digital Scale Owner’s 22
$ (SCALE CAPACITY)
The display will show $ and the current setting. If the value displayed is acceptable,
press the key. Otherwise, press the
×
or
Ø
keys to select a new value and then press
the key to save it and proceed to the next prompt.
This is the maximum allowed weight of the scale. It should be set to 600.0.
)0; (INTERVAL SETTING)`
The display will show )0; and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to select a new value and then press
the key to save it and proceed to the next prompt.
This is the minimum graduation of the scale. Allowable settings are 1, 2, or 5.
3 (DECIMAL POINT POSITION)`
The display will show 3 and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to select a new value and then press
the
key to save it and proceed to the next prompt.
This is the decimal point precision of the scale. Allowable settings are 0, 1, 2, or 3.
0 = XXXXX 1= XXXX.X 2 = XXX.XX 3 = XX.XXX
*% (SAMPLE RATE)
The display will show *9 and the current setting. If the value displayed is acceptable, press
the
key. Otherwise, press the
×
or
Ø
keys to select a new value and then press the
key to save it and proceed to the next prompt.
This is the number of times per second the load cell is sampled. Allowable values are a
minimum of 1 sample per second to a maximum of 10 samples per second in one sample
per second intervals.
)!& (DIGITAL FILTER MODE)
The display will show )!; and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to select a new value and then press
the
key to save it and proceed to the next prompt.
This is the amount of digital filtering applied to the scale. Allowable settings are 0, 1, 2, or 3.
0 = Off 1 = Minimal 2 = Moderate 3 = Maximum
00#& (MOTION RANGE)
The display will show 00#; and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to select a new value and then press
the
key to save it and proceed to the next prompt.
The motion range is the number of divisions of change permitted before indicating unstable
(the STABLE annunciator turns off). Allowable values are 1 to 10.