EasyManua.ls Logo

DeVilbiss IntelliPAP DV54 - DV5 HH Humidifier General Information; DV5 HH Safety Safeguards; DV5 HH Travel and DC Power; DV5 HH Product Description

DeVilbiss IntelliPAP DV54
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
LT-2 027
EN
11. General Informat ion DV 5H H
A. SAFEGUARDS
When servicing electrical products, basic safety precautions
should always be followed. Important safety information in this
manual is highlighted by the following terms.
DANGER: Urgent safety information for hazards that
will cause serious injury or death.
WARNING: Important safety information for hazards
that might cause serious injury.
CAUTION: Information for preventing damage to the
product.
NOTE: Information to which you should pay
special attention.
PLEASE READ ALL INSTRUCTIONS BEFORE
SERVICING THIS DEVICE.
DANGER!
Electric Shock Hazard – Do not use while bathing.
Electric Shock Hazard – Do not immerse this
device into water or any other liquid.
Electric Shock Hazard – Do not attempt to open or
remove the enclosure. There are no user-serviceable
internal components. If service is required, return
the product to your home care provider. Opening or
tampering with the product will void the warranty.
WARNING!
Use the Heated Humidifi er system only for its
intended use as described in this manual.
Be sure to read and understand all safety instruc-
tions supplied with your IntelliPAP or SleepCube fl ow
generator device.
Use only accessories recommended by DeVilbiss.
Always remove the water chamber from the humidi-
er cradle for fi lling.
Use only tubing recommended or supplied with the
ow generator.
If water has been spilled onto the cradle or it has
been submerged into water, unplug the power cord
from the power source immediately. Allow the de-
vice to completely dry before use.
If the device has been dropped, refer to the trouble-
shooting guide for instructions. Contact your homec-
are provider for evaluation of the device.
Never block the air openings of the cradle or chamber.
Do not insert any objects into any openings or tubes.
For proper operation of the humidifi er, the device
must be placed on a fl at, level surface.
Never place the system on a soft surface such as a
bed or couch during operation.
This heated humidifi er is intended for single-patient
use only.
B. TRAVEL
The DeVilbiss DV5HH humidifi er system
Automatically accepts line voltages of 100-240 V, 50/60 Hz
Needs power cord appropriate for area
o USA DV51D-606
o Europe, except UK DV51D-607
o UK DV51D-608
o Australia DV51D-609
o Set of 3 (UK, Europe, Australia) DV51D-611
C. DC POWER
The DeVilbiss DV5HH humidifi er system
Will NOT operate, if the DC power is coming
directly from a 12V DC power source.
Will operate, if the DC power coming to the PAP
passes through an inverter so that AC power is
delivered to the system.
See DC Power information under the DV54 section
of this manual
D. PRODUCT DESCRIPTION
The DV5HH Standard Heated Humidi er System is intended to
warm and add moisture to the pressurized air supplied to the pa-
tient during the treatment of Obstructive Sleep Apnea (OSA). The
addition of heated humidi cation to the air relieves dryness and
irritation to the patient’s airway during OSA therapy. The heated
humidifi er is used in conjunction with the DV5X series PAP devices
and consists of a humidifi er chamber and humidi er heater.
The humidifi er chamber holds enough water for a minimum
of 8 hours of operation, with the heating unit at the maximum
temperature setting at ambient conditions of 23°C and 25%
R.H. with a system fl ow rate of 60 L/Min. The chamber slides
out of the unit for cleaning and fi lling, without adjusting tubing
connections. The chamber’s halves separate for easy cleaning
but, when assembled, maintain a leak free seal to a maximum
operating pressure of 30 cmH2O.
Heat transferred from the humidifi er heater to the humidifi er
chamber raises the effi ciency of the DV5HH system. The pa-
tient controls the heater plate temperature using the up/down
arrow keypad keys on the attached PAP device to provide the
appropriate level of humidifi cation depending upon environ-
mental conditions and individual preference.