Device Symptom Potential Problem Corrective Action
6.2.9 Handset Failure
alarm displayed on
screen.
The handset, in general, is
not communicating with
the device.
Ensure handset is securely attached.
The batteries are impaired
or at the end of their life
cycle.
Replace with a newly charged handset in
order to continue. Use the charging cradle to
determine the handset status. If the charging
bars illuminate solid red or if no illumination
occurs, the handset will require authorized
service. See Section 7.0 Maintenance and
Repair.
A different problem exists
if all handsets appear
fully charged.
If this condition persists, authorized service is
required. See Section 7.0 Maintenance and
Repair.
6.2.10 The handset did
not seem to be working
when the trigger was
pulled.
The trigger is not
communicating with the
device.
If this condition persists, authorized service is
required. See Section 7.0 Maintenance and
Repair.
6.2.11 NOMAD Pro
2 automatically shuts
down.
NOMAD Pro 2 times out
after about three minutes
of inactivity.
Manually turn on NOMAD Pro 2 when you
are ready to use the device.
A different problem exists
if shut-down occurred
during regular activity.
If this condition persists, authorized service is
required. See Section 7.0 Maintenance and
Repair.
6.2.12 Ready state
terminated before an
exposure started.
The Ready state expires
because the X-ray
exposure is not initiated
within 15 seconds of the
start of the Ready state.
Double-check the exposure time setting and
re-enable when operator and patient are
again properly situated.