DS-685L/F Series Service Manual
Page No. 15
3. Error Message
The following error message will appear when a incorrect operation is performed.
Message Reasons Appropriate
O F
When displayed weight exceeded
capacity+9d, or something is on the
platter when power on.
Remove the item on the platter.
U F
When displayed minus weight < -9d
Re-Zero or On/Off again, if still no
good please check the Load Cell or
A/D board are problem or not.
E R R O R
When calibration operation is not
correct, or A/D starting check error.
Repeat calibration operation, if still
no good please check the load cell
or A/D board are problem or not.
# # # # #
(*Note)
Scale is not steady when power on
Place scale on firm, flat base. If still
no good please try to recalibrate it.
totAL FULL
When current Total overflow.
Clear current Total, or check the
memory chip.
G.t. FULL
When Grand Total overflow.
Clear Grand Total, or check the
memory chip.
Note: “ # # # # # ” - All segment display.
If you still cannot remove the trouble what you met according to above information, please
contact to our Technical Service Dept. as cover page indicated.
All manuals and user guides at all-guides.com

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Digi DS-685L Series and is the answer not in the manual?

Digi DS-685L Series Specifications

General IconGeneral
BrandDigi
ModelDS-685L Series
CategoryScales
LanguageEnglish

Related product manuals