Product description
www.dilo-gmbh.com | 0200.bae
Settings:
When pressing the “Settings“ screen selection key the settings screen can be paged through and
modifications can be carried out.
Note for hysteresis:
Input of final vacuum: e. g. 5 Torr
Input of hysteresis: e. g. 4 Torr
The suction pump starts the recovery process after having reached the final vacuum of 1 Torr (difference
final vacuum to hysteresis). Finally a vacuum test is carried out. If the pressure does not exceed the final
vacuum the function is completed automatically.
Note:
the entered final vacuum must be higher than the hysteresis entered!