DFW series indicators DFW_03_12.07_EN_U
49
- It is not possible to just cancel a single weigh.
- The initialisation is possible only with a non approved instrument.
SERIAL COMMANDS
Besides the commands described in the section “FORMAT OF THE SERIAL COMMANDS”, TECH.MAN.REF., in this
functioning mode also the commands below are available:
WEIGH STORAGE
Command
[II]PID<CRLF> or <ESC>[II]PID<STX>
[II]PIDD<CRLF> or <ESC>[II]PIDD<STX>
in which: [II]: 485 address
<ESC>: 27 ascii decimal character
<STX>: 2 ascii decimal character
Instrument response to the [II]PID<CRLF> command:
[II]PIDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO) <CRLF>
Instrument response to the <ESC>[II]PID<STX> command
:
<ESC>[II]PIDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO)<STX>
Instrument response to the [II]PIDD<CRLF> command:
[II]PIDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO),(dd/mm/yybbhh:mm:ss|"NO DATE TIME")<CRLF>
Instrument response to the <ESC>[II]PID<STX> command
:
<ESC>[II]PIDSS,B,LLLLLLLLLLUU,YYTTTTTTTTTTUU,(ID | NO),(dd/mm/yybbhh:mm:ss|"NO DATE TIME") <STX>
In which:[II] 485 address (only when transmitting in 485 mode)
SS OL" (weight in overload) or "UL" (weight in underload) or "ST" (stable weight) or "US" (unstable
weight) or "TL" (TILT input closed).
B scale number (always 1)
LLLLLLLLLL: gross weight on 10 digits
UU: unit of measure
YY: 2 spaces in the case of null tare or semiautomatic tare, “PT” in case of manual tare
TTTTTTTTTT: tare on 10 digits
ID XXXX-YYYYYY in which: XXXXX is the rewriting number (5 digits, from 00000 to 00255) and
YYYYYY is the weigh number (6 digits, from 000000 t0 131072).
dd/mm/yy Date in the "dd/mm/yy" format (only with PIDD command).
bb 2 space characters, 32 decimal ascii character (only with PIDD command).
hh:mm:ss Time in the "hh:mm:ss" format (only with PIDD command).
In the case in which the gross weight is negative or unstable, the weight is transmitted but not the ID; “NO” is in its place. In
these cases there is no storage in the alibi memory.
In the case in which the date/time is not detected or set, the weight is transmitted but not the date and time; “NO DATE
TIME” is in its place.
WEIGH READING
Command:
[II]ALRDXXXXX-YYYYYY <CR o CRLF>
In which: [II] 485 address (only when transmitting in 485 mode)
XXXXX rewriting number (from 00000 to 00255)
YYYYYY weigh number (from 000000 to 131072)