B-50
1-16. Spindle Coolant Flow Check
Procedures to check the flow rate of spindle coolant water
Step No.
Procedure
(Continued from the previous section)
1
Remove the cover shown in the figure below at the left side of the machine.
Remove this cover
Spindle coolant flowmeter
Left side
2
Check to make sure that the flow rate of spindle coolant water is 1.5 L/min or higher.
If the flow rate of spindle coolant water is less than 1.5 L/min, check the flow rate of
supply water at the plant facility side.
3
Install the cover removed in step 1.