Operating Manual•Mode D'Emploi
Bediehnungs Handbuch •Manual de Operación
Manual de operação
Fluid Flow Range:
Débit d'eau:
Durch ussmenge:
Caudal de trabajo:
Vazão Operativa:
0.15 gpm to 30 gpm
0,57 l/mn to 114 l/mn
Injection Range
Dosage:
Dosierung:
Dosi cación:
Injeção:
0.025% to 5%
1:4000 to 1:20
Operating Pressure:
Pression:
Druck:
Presión operativa:
Pressão operativa:
5 to 100 psi
0,34 to 6,9 bar
*Speci cations vary by model.
*Les données techniques varient
selon les modèles.
* Technische Daten sind je nach
Modell unterschiedlich.
* Características técnicas varían
según modelo.
* Características técnicas variam
conforme o modelo.
Model 0.3%
Model 0.3% PAA
Model 2.5%
Model 2.5% WSP
Model 5%

Table of Contents

Questions and Answers

  • J
    Joseph BoyerAug 2, 2025
    Why is my Dosmatic Superdos 30 Laboratory Equipment not making a clicking sound?
    • S
      Scott MosleyAug 2, 2025
      If your Dosmatic Laboratory Equipment isn't clicking, several factors could be at play. If fluid is not flowing through the unit, ensure the red plugs at the inlet, outlet, and suction tube fitting openings are removed and that the unit isn't installed backward—the arrow should align with the fluid flow. If it still doesn't click, contact Dosmatic Customer Service. If fluid is flowing, the fluid rate might be outside the injector's rated service flow, or the injector may need to be submerged in room temperature fluid for 24 hours to reabsorb fluid. High operating pressure could also be the cause; consider installing a pressure reducer valve. Check that the On/Off lever is in the ON position and the By-Pass Valve (A) is closed.
  • D
    Dustin PowersAug 9, 2025
    What to do if my Dosmatic Superdos 30 Laboratory Equipment is clicking but not suctioning solution?
    • A
      Ashley AllenAug 9, 2025
      If your Dosmatic Laboratory Equipment clicks but doesn't suction solution, it could be due to several issues. The cylinders #7 or #68 might be worn and need replacement. For WSP models, the dosage piston/gasket #44 (or #51 for 0.3% models) could be worn or incorrectly installed; replace it, ensuring the flared-end faces up during installation. Also, check the O-ring retainer #15 for correct installation, and inspect the O-ring seat #14 or dosage piston/gasket for damage, replacing as necessary. A worn or loose #17 O-ring, a cracked, leaking, or clogged suction tube #25 or fitting #11, or a leaking check valve #13 could also be the cause, requiring replacement or cleaning.
  • D
    Diane BrandtAug 15, 2025
    How to fix under injecting in Dosmatic Laboratory Equipment?
    • C
      christopherruizAug 15, 2025
      If your Dosmatic Laboratory Equipment is clicking but under-injecting, the #44 (or #51 for 0.3% model) Dosage Piston/Gasket (WSP model) or the #7 (or #68 - 0.3%) Inner Cylinder might be worn and need replacement. If the unit operates well at high flow but not at low flow, consider replacing the #17 O-ring. Also, the Main Piston Assembly #9, Cover #1, or main body #40 could be worn or scored, requiring replacement or cleaning of the fluid filter.
  • M
    Matthew SmithAug 20, 2025
    What to do if Dosmatic Superdos 30 Laboratory Equipment makes no clicking sound?
    • J
      Jason SmithAug 20, 2025
      If your Dosmatic Laboratory Equipment isn't clicking, it could be due to a worn Main Piston Assembly #9, or worn or scored Cover #1 or main body #40, which would require replacement or cleaning of the fluid filter. Ensure the On/Off lever is in the ON position, and the By-Pass Valve (A) is closed. A dirty or plugged inlet filter may also be the culprit; ensure the mesh size is correct and clean the filter. Finally, #17 might be worn or not seated properly, requiring reseating or replacement.
  • A
    alicialewisAug 27, 2025
    Why is fluid refilling the solution tank in my Dosmatic Superdos 30 Laboratory Equipment?
    • J
      John WilliamsAug 27, 2025
      If fluid is refilling the solution tank in your Dosmatic Laboratory Equipment, the check valve #13 might be leaking. Inspect the seat area on the suction tube fitting #11, ensuring the check valve and seal fit loosely within. Clean the seal and inside fitting to remove any debris. Alternatively, the washer seal on #13 may be swollen or damaged by chemical attack, in which case you should replace it with a new check valve assembly.

Summary

Operating Principle

Accurate and Reliable

Describes how the injector operates using fluid pressure.

Maintenance and Warranty

For Your Records

Space to record injector serial number and date of purchase.

Important: For Product Registration

Encourages product registration via the Dosmatic website.

Package Contents

Specifications

Details technical data like flow rate, pressure, and materials.

Safety Precautions

Warranty Compliance

Details on adhering to warranty terms and conditions.

General Tips

Advice for proper usage and extended injector life.

Operations

Bypass Operation

How to temporarily stop injection using the ON/OFF feature.

Installation and Start-up

Suggested Installation Diagram

Illustrates recommended components and their placement for setup.

Maintenance

Remote Injecting

Explains the purpose and recommendation for remote injection kits.

Routine Maintenance Instructions

Troubleshooting

Injector Repair Parts

Lower End & Wear Parts Kits 0.3% PAA (Rotating)

Lists wear parts kits for the 0.3% PAA (Rotating) model.

Lower End & Wear Parts Kits 0.3% (Rotating)

Lists wear parts kits for the 0.3% (Rotating) model.

Lower End & Wear Parts Kits 2.5%

Lists wear parts kits for the 2.5% model.

Lower End & Wear Parts Kits 2.5% WSP

Lists wear parts kits for the 2.5% WSP model.

Lower End & Wear Parts Kits 5%

Lists wear parts kits for the 5% model.

Warranty

To Return an Injector for Warranty or Non-Warranty Repair

Instructions for sending the unit for service or evaluation.