32
Instructions for Use Air Compressor
Dew point depression at operating pressure 5 °C (41 °F) under ambient temperature at 30 L/min
and max. ambient temperature 40 °C (104 °F).
15 °C (59 °F) under ambient temperature at 15 L/min
and max. ambient temperature 40 °C (104 °F)
Air quality Filtered and oil-free compressed air
Filter m
Compressed air outlet Coupling with check valve
High temperature indicator visual (orange LED) and acoustic (continuous buzzer)
Standby – pressure monitoring
Compressor – compressed air ON <2.7
±0.2 bar (<40±3 psi)
Compressor – compressed air OFF >3.2
±0.2 bar (>46±3 psi)
Operating characteristics
Connection for central gas supply system NIST or DISS connection
Compressor Nominal voltage Frequency Standby mode Fuses (T1, T2)
5x20mm
Current
consumption
[V] [Hz] IEC 127-2/V [A]
8413417 100 50 yes T8H250V 5.4
8413894 100 60 yes T8H250V 5.4
8413418 110 50 yes T8H250V 4.8
8413899 110 60 yes T8H250V 4.8
8413419 115-120 60 yes T8H250V 4.7
8413421 115-120 60 yes T8H250V 4.7
8413422 127 60 yes T8H250V 4.2
8413423 127 50 yes T8H250V 4.2
8413892 220-230 60 yes T4H250V 2.4
8413897 220-240 50 yes T4H250V 2.2
8413900 220-240 50 no T4H250V 2.2
Protection class I
Electromagnetic compatibility EMC Tested acc. EN 60601-1-2: 2004 and 2007
Sound pressure level at 50 Hz operation approx. 49 dB(A) at maximum continuous flow
Sound pressure level at 60 Hz operation approx. 52 dB(A) at maximum continuous flow
Measured at the front at a distance of 1 m (3.28 ft)
and a height of 1 m (3.28 ft).