Instructions for Use Air Compressor 31
English / US English
Disposal
For countries subject to the EU directive
2002/96/EC
This device is subject to EU Directive 2002/96/EC
(WEEE). In order to comply with its registration
according to this directive, it may not be disposed
of at municipal collection points for waste electrical
and electronic equipment. Dräger has authorized a
company to collect and dispose of this device. To
initiate take-back or for further information, visit us
on the Internet at www.draeger.com and navigate
on the DrägerService area where you will find a link
to "WEEE". If you have no access to our website,
contact your local Dräger Medical Organization.
Disposal of the medical device
When disposing of the medical device:
Consult the relevant waste disposal company
for appropriate disposal.
Observe the applicable legal requirements.
Technical Data
Ambient conditions
Operation
Temperature 10 to 40 °C (50 to 104 °F)
Air pressure 700 to 1060 hPa
Relative humidity 10 to 90 %, no condensation
Storage
Temperature –20 to 60 °C (–4 to 140 °F)
Air pressure 500 to 1060 hPa
Relative humidity 5 to 95 %, no condensation
Performance characteristics
Operating pressure 4
+1.0 bar (58+14.5 psi)
When the standby mode is used, the output pressure
is at least equal to the input pressure.
Continuous flow minimum 30 L/min at 3.0 bar (44 psi)
The specified continuous flow always relates to the
respective ambient pressure present.
Peak flow 180 L/min for max. 0.8 seconds at a total of max.
30 L/min