12. Error Messages and Operation References
After being switched on, as well as during the measurement process, the MSI EM200-E
checks the function of all measuring channels (see 3.3). Error messages are shown before
the selection of functions or during normal function.
12.1 Error Messages
12.1.1 Error Messages Regarding Sensors
Error Message Error Cause
Remedy
CO-Sensor
Probe has been in flue during zero calibration
Sensor defect
Calibrate again with fresh air
Service
NO-Sensor
NO-Sensor has been without current
Probe has been in flue during zero calibration
Sensor defect
check function
see CO-sensor
NO2-Sensor
Probe has been in flue during zero calibration
Sensor defect
Calibrate again with fresh air
Service
Pressure Sensor Sensor defect
Service
12.1.2 General Error Messages
Error Message
Error Reason; Remedy
System temperature operating temperature range exceeded;
vary temperature, service
Battery temperature
operating temperature range exceeded;
vary temperature, service
Battery voltage voltage not between 4.4 V and 6.5 V; replace battery (service)
Battery current battery current for charging or operating too high; service
Battery error battery manager data error; charge battery
Set clock clock lost data; charge battery, set clock
Options
data record options error; factory service
Settings
wrong settings; check and change settings
Calibration data
error in calibration data record; service
Data memory
error in memory data record; service
Next service
next service data error; service
System configuration
system configuration error, service