Other mode steps:
002 Check medical gas supply:
or
003 Test switch for ventilation modes:
004 Check settings of rotary knob for minute volume
alarm limits
005 Sound source test, lamp test, display test
006 Check settings of rotary knobs for "f", "VT",
"Tinsp", "Pmax"
007 Switching ON / OFF high rate alarm
or
Exit Customer Service Mode:
● On / off switch to 0 (OFF)
INFO TEST ON
RESET TEST OFF
F = XX.XX VT = X.XX
Ti = X.XX P = XX.X
MV-low = XX.XX
MV-high = XX.XX
Pos. IPPV/SIPPV
Pos. SIMV/CPAP
Tachypnea alarm
ACTIVE
Tachypnea alarm
INACTIVE
49
Pressure high
Pressure low
Customer Service Mode
Operating Instructions MicroVent