Section 5 - Operation
Respiratory Volume Monitoring
Monitor Controls
You use the minute volume alarm keys to set minute volume alarm
limits and to turn the volume alarms on and off. These keys are in the
bottom row of the key bank to the right of the right-hand display screen.
INCREMENT
K
EY
BPM
TID VOL
MIN VOL
5.6100.58
OP20031
LO
MINUTE VOLUME ALARMS
VOLUME
ALARMS
DISABLE
VOLUME
ALARMS
DISABLE KEY
DECREMENT
KE
Y
3.0
L
OW MINUTE VOLUME
A
LARM LIMIT
A
DJUSTMENT KEY
Fixed alarms are provided for low tidal volume, low minute volume, and
reverse flow through the sensor. Apnea volume alarms are generated at
15 seconds (Caution) and 30 seconds (Warning) if the respiratory volume
monitor does not sense a valid breath.
The Narkomed 2Bs volume alarms are automatically enabled when the
ventilator power switch is turned to the ON position. A disconnected
sensor causes a sensor disconnect alarm.
5-11-3