Section 5 - Operation
AV2+ Anesthesia Ventilator
I:E RATIO CONTROL
I:E RATIO DISPLAY
FREQUENCY
CONTROL
FREQUENCY
DISPLAY
INSPIRATORY FLOW GAUGE
INSPIRATORY
FLOW CONTROL
VENTILATOR
ON-OFF
CONTROL
TIDAL VOLUME
CONTROL
PRESSURE
LIMIT CONTROL
TIDAL
VOLUME
SETTING INDICATOR
BELLOWS CANISTER
BREATHING CIRCUIT
CONNECTOR
OP91018c
FREQUENCY
/min
I:E RATIO
INSPIRATORY
FLOW
VENTILATOR
ON
FAULT
1400
1200
1000
800
600
400
200
PRESET TIDAL VOLUME (ml)
EXTENDED RANGE
ACCESS
EXTENDED
RANGE
AV2+
INSPIRATORY PRESSURE LIMIT
cmH
2
O
TIDAL VOLUME
PUSH TO TURN
Activating the
Ventilator
The ventilator can be activated by using the ventilator on/off control, or,
optionally, by using the lever on the manual/automatic selector valve.
The anesthesia machine’s SYSTEM POWER switch must be set to ON in
order for the ventilator to function.
NOTE: The selector valve lever can activate the ventilator only when
the interface cable is connected between the manual/automatic
selector valve and the SELECTOR fitting on the breathing
system sensor interface panel (see following illustration);
when the interface cable is disconnected, the position of the
selector lever has no effect on the ventilator.
5-8-2