Section 7
Specifications
Vaporizers
(Vapor 19.1)
Temperature Range ................................ +15–35°C
(at normal atmospheric
pressure of 760 mmHg)
Flow Range .................................... 0.25-15 l/min
Maximum Pressure Load ........................... 150mmHg
(above atmospheric)
Maximum Angle of Inclination ............................ 45°
Weight ................................ Approximately 7.5 kg
The following values refer to individual concentration settings when
operated with a continuous flow of air in the range 0.25–15 l/m,
temperature at 22°C, and normal atmospheric pressure (760 mmHg).
Halothane
Adjustment range .................................0.2–5 vol%
Accuracy ..................... ±0.15% concentration (volume) or
± 15% (whichever is higher)
Enflurane
Adjustment range .................................0.3–7 vol%
Accuracy ...................... ±0.2% concentration (volume) or
± 20% (whichever is higher) or
+ 20%/- 30% with flow settings
6.0–15 l/min and handwheel settings
higher than 5.0% volume concentration
Isoflurane
Adjustment range .................................0.2–5 vol%
Accuracy ..................... ±0.15% concentration (volume) or
± 15% (whichever is higher)
Sevoflurane
Adjustment range .................................0.3–8 vol%
Accuracy ...................... ±0.2% concentration (volume) or
± 20% (whichever is higher)
Ventilator
Frequency .................. 1–99, ±1 BPM (in 1 BPM increments)
I:E ratio ...... Standard range: 1:1–1:4.5, ±0.1 (in increments of 0.5);
Extended range: 4:1, 3:1, 2:1
Inspiratory flow ......................10–100 l/min (uncalibrated)
Tidal volume .............................50–1500 ml, ±100 ml
Pressure limit control adjustment range ............. 15–120 cmH
2
0
Absorber System
Inspiratory Valve
Mounting ring nut size ...............................M35x1
Hose terminal .................................. 22mmmale
Expiratory Valve
Mounting ring nut size ...............................M33x1
Hose terminal .................................. 22mmmale
7-3