Section 2
General Description
Power Supply
System
The Narkomed 2B is equipped with a central power supply for the
ventilator, alarm system, and monitoring system. When in use, the
Narkomed 2B must be plugged into an AC outlet.
Convenience
Receptacles
The Narkomed 2B is equipped with four convenience receptacles,
mounted on the under side of the rear of the monitor bank. (Machines
with the 240 VAC power supply are not equippped with convenience
receptacles). The receptacles are active whenever the Narkomed 2B is
plugged into an outlet, whether or not the machine is turned on.
The total current for devices plugged into the receptacles must not
exceed 5 amps.A5ampcircuit breaker protects the convenience
receptacle circuit. This circuit also incorporates an EMI filter, which
minimizes interference to the anesthesia machine from devices plugged
into the convenience receptacles.
CAUTION: Devices plugged into the convenience receptacles
contribute to the anesthesia systems total leakage current.
The total leakage current (for system and external devices
combined) must not exceed 100 microamps.