Section 5 - Operation
Gas Delivery System
OP14700
DISS NUT
OPTION
N2O DISS NUT
GAS FITTING
N2O DISS BODY
GAS FITTING
N2O NIST
GAS FITTING
DISS BODY
OPTION
NIST
OPTION
O2 DISS BOD
Y
GAS FITTIN
G
AIR DISS NUT
GAS FITTING
O2 DISS NUT
GAS FITTING
AIR NIST
GAS FITTING
O2 NIST
GAS FITTING
AIR DISS BODY
GAS FITTING
Connecting the
Gas Cylinders
When attaching a cylinder, make sure that only one washer is installed
between the cylinder and the yoke gas inlet. Using multiple washers
may compromise the pin-indexed safety system. Be sure to verify the
integrity of both index pins whenever you install a new cylinder.
To connect a gas cylinder to its yoke:
1. Place a new washer on the seat of the yoke gas inlet connection.
WARNING: Use only one cylinder washer per yoke. Using more than
one washer could cause leakage of the cylinder gas and
compromise the pin indexing system.