Section 5 - Operation
Breathing Pressure Monitoring
Right-Hand ScreenNumerical values for PEEP (positive end
expiratory pressure), mean, and peak breathing pressure are shown
on the right-hand screen.
MEANPEEP PEAK
44112
OP20027
HI
PRESSURE ALARMS
APNEA
ALARM
DISABLE
PEEP
BREATHING
PRESSURE
MEAN
BREATHING
PRESSURE
PEAK BREATHING
PRESSURE
12
50
HIGH PRESSURE ALARM LIMIT
THRESHOLD PRESSURE
ALARM LIMIT
The PEEP pressure is the pressure at the end of exhalation.
The mean pressure represents the average of all the instantaneous
pressure values recorded during each breath.
The peak pressure is the highest instantaneous pressure value for
each breath.
To the right of the peak variable, in small type, are two alarm limits.
The upper number is the pressure high alarm limit; the lower
number is the threshold pressure alarm limit.
5-12-5