Section 5 - Operation
AV2+ Anesthesia Ventilator
PPRROOBBLLEEMM PPOOSSSSIIBBLLEECCAAUUSSEE RREEMMEEDDYY
Ventilator does not
operate; yellow
FAULT LED on
ventilator lights
Selector switch on the
absorber is in the
BAG position
Turn the selector
switch on the
absorber to the
AUTO position.
Bellows wont
compress during
inspiration
Absorber manual/
automatic selector
valve in BAG position
Place selector valve
in AUTO position.
Inspiratory flow
control setting on
ventilator too low
Increase inspiratory
flow control setting.
Frequency too high Decrease frequency.
Inspiratory pressure
limit setting on
ventilator too low
Increase inspiratory
pressure limit
setting.
5-8-7