Troubleshooting
Dräger Polytron 8000 27
9 Troubleshooting
9.1 Fault reference
Fault number Cause Remedy
#001
#003
#004
#005
#011 - #014
#020 - #024
#043
#060
#067
Serious device fault,
various causes.
Have the device checked by DrägerService
®1
.
#002
#025 - #027
#030 - #034
#041
#042
#044
#050
#052
#081
#085
Serious data error in the device,
various causes.
Reset the device to factory settings (see chapter 8.2.6 on
page 21).
If this fault occurs again:
Have the device checked by DrägerService.
#010 4-20 mA interface cable not connected. Disconnect the power supply, connect the 4-20 mA interface
and reconnect the power supply.
For operation without controller, connect terminal 3 with
terminal 4 (see chapter 3.5.1 on page 8).
#045 Device not detecting any sensor. Check contacts.
If this fault occurs again:
Have the device checked by DrägerService.
#051
#054
Faulty zero-point. Perform zero-point calibration (see chapter 5.1.1 on page 15).
#055 Faulty span calibration was performed. Perform new span calibration (see chapter 5.1.2 on page 16).
#076
#080 - #084
Device fault. Check contacts.
If this fault occurs again:
Have the device checked by DrägerService.
#105 Device fault. Have the device checked by DrägerService.
#137 Device fault. Restart.
If this fault occurs again:
Have device checked by DrägerService.
1 DrägerService
®
is a registered trademark of Dräger.