32 Dräger Polytron 8000
Technical data
12.1 Torques
(The torques are valid for the aluminum and stainless steel
316 variants)
If an optional e-Box is being used, fasten the device securely
to the e-Box with 4 screws and a tightening torque of 8 Nm.
12.2 Tightening torque and cable size for field
wiring terminals
12.3 Factory settings
12.3.1 Alterable settings for the Polytron 8000
Part Torque Lb. In. Torque Nm
Housing cover min. 266 min. 30
Sensors min. 266 min. 30
Plug min. 266 min. 30
Cable bushing min. 443 min. 50
Device to e-Box 71 8
Terminals Torque Lb. In.
Cable size in
AWG
Cable size in
mm
2
Power
supply and
signal
4.4 – 7.0
(0.5 – 0.8 Nm)
24 – 12 0.2 – 2.5
Relay
4.4 – 7.0
(0.5 – 0.8 Nm)
20 – 12 0.5 – 2.5
Menu Standard setting
A1 Alarm
Depending on the sensor
A2 Alarm
A1 Relay alarm:
Direction
Rising
(falling with DrägerSensor O
2
)
A2 Relay Alarm:
Direction
Rising
A1 latching mode Non-latching
A2 latching mode Latching
A1 Acknowledgement mode Acknowledgeable
A2 Acknowledgement mode Not acknowledgeable
A1 enrg / wired in e-Box Energized / NO
A2 enrg / wired in e-Box Energized / NO
Calibration password _ _ _1
LCD setting On
Configuration password _ _ _2
SIL status Off
Language English
Fault current 1.2 mA
Warning current 3.0 mA
Warning Off
Maintenance current 3.4 mA
HART address 0
Auto-cal. Off
Calibration interval Depending on sensor
Function key Faults

Table of Contents

Other manuals for Dräger Polytron 8000

Related product manuals