5327.400 Vapor 2000 05/99 Function Description Page 16
For internal use only. Copyright reserved.
GBF5327400T02.fm 28.08.01
4 Differences between Vapor 19.n and Vapor 2000
Vapor 19.n Vapor 2000
Filling volume approx. 140 mL with
damp wick
approx. 300 mL with damp
wick
Permitted tilt Up to 45° Any in “T” setting and up to
30° in operation
Logistics 70 variants
70 part numbers
4 part numbers for all
variants
Temperature range 15 to 35 °C 10 to 40 °C
Flow control range 0.2 to 5 vol.%
1
2
1. Halothane
2. Isoflurane
H 1 and I 2 0.2 to 6 vol.%
0.2 to 7 vol.%
3
3. Enflurane
E 3 and S 4 0.2 to 8 vol.%
0.2 to 8 vol.%
4
4. Sevoflurane
Draining of anaesthetic prior to
transportation
Yes No
Low-flow suitability Good Very good
Vaporizer can be switched on
without locking plug-in system
Possible Not possible
Control dial cap Neutral Coloured
Stability 3-leg Flat, so very good
Sightglass screen Not provided Fitted
Vaporizer removable when
switched on
Yes No