This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 28
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
5.1.2 Test plug-in adapter locking lever
• Remove the vaporizer from the anaesthetic machine.
• Check that the plug-in adapter locking lever jumps back automatically.
• Check that the locking pin on the locking lever is undamaged (firm and
straight).
• Check that both valve operating pins are present.
• Check that the sealing surfaces are undamaged.
5.2 Plug-in adapter S-2000
(Selectatec, white version)
5.2.1 Test plug-in adapter locking lever
• Remove the vaporizer from the anaesthetic machine.
• Check that the plug-in adapter locking lever jumps back automatically.
• Check that the locking pin on the locking lever is undamaged (firm and
straight).
• Check that the interlock pins move freely.
• Check that both valve operating pins are present.
• Check that the sealing surfaces are undamaged.
s for sevoflurane
Plug-in adapter DW 2000 Code ID: