This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 26
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
4.7 Test keyed filler adapter
• Check that the relevant keyed filler adapter has no defects:
• Check that the keyed filler adapter is coded to the specified anaesthetic
at both ends.
• Check that the relevant filling system is fitted with the necessary ball
valve.
• Check that the O-ring in the cylinder port is present and undamaged.
• Check that the keyed filler adapter fits in the intended filling opening.
Keyed filler
adapter:
Part number
h for halothane M 30288
e for enflurane M 30289
i for isoflurane M 30290
s for sevoflurane M 31930