This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 26
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
4.7 Test keyed filler adapter
Check that the relevant keyed filler adapter has no defects:
Check that the keyed filler adapter is coded to the specified anaesthetic
at both ends.
Check that the relevant filling system is fitted with the necessary ball
valve.
Check that the O-ring in the cylinder port is present and undamaged.
Check that the keyed filler adapter fits in the intended filling opening.
Keyed filler
adapter:
Part number
h for halothane M 30288
e for enflurane M 30289
i for isoflurane M 30290
s for sevoflurane M 31930