This test list does not replace the inspections and servicing to be carried out by DrägerMedical AG & Co. KGaA
5327.400 Vapor 2000 04.99 Test list Page 30
GBK5327400TL1.fm 19.01.00
Copyright reserved. Reproduction permitted only for non-commercial purposes.
7 Test ISO connector for vaporizers
• Check that the male cone at the vaporizer inlet and the female cone at
the vaporizer outlet are connected.
8 Leak test
8.1 Test vaporizer on associated anaesthetic machine
• Lock the lock catch or cap.
• Switch on the vaporizer (> 0.2 vol.%).
• Depressurise the AIR system and the N
2
O system.
• Close all flow control valves.
• Connect the mixed gas socket to the pressure gauge.
•Set the O
2
flow control valve such that a pressure of 30 mbar is
indicated on the pressure gauge.
• Check that the gas flow is < 0.12 L/min.