7 Test ISO connector for vaporizers
Check that the male cone at the vaporizer inlet and the female cone at
the vaporizer outlet are connected.
8 Leak test
8.1 Test vaporizer on associated anaesthetic machine
Lock the lock catch or cap.
Switch on the vaporizer (> 0.2 vol.%).
Depressurise the AIR system and the N
2
O system.
Close all flow control valves.
Connect the mixed gas socket to the pressure gauge.
•Set the O
2
flow control valve such that a pressure of 30 mbar is
indicated on the pressure gauge.
Check that the gas flow is < 0.12 L/min.