Introduction
8 Vista 120 Gateway User Manual
Introduction
Intended Use
Vista 120 Gateway (hereafter called “Gateway”) is intended to receive physiological data collected by
patient monitors (hereafter called “monitors”), send the data to external systems (e.g. HIS) after the data
are converted to HL7 messages, and transmit the control information sent by external systems to
monitors.
Scope of Application
Gateway is applied to load ECG, RESP, NIBP, SpO2, PR, TEMP, IBP, CO, CO2, AG, BIS, VENT, and AM
data collected by the manufacturer’s monitors (e.g. Vista120, Vista120 S) to external systems in hospital.
NOTE
Please read this chapter carefully before you use the software and ask professionals authorized by the
manufacturer to connect and install Gateway.
NOTE
All the descriptions or pictures in this manual are based on Windows Server 2016 Standard 64 bit
system. Operation descriptions or pictures in other systems may differ from that in Windows Server 2016
Standard 64 bit.
NOTE
While a monitor measures real-time NIBP and at the same time, Gateway sends a periodic R01
message to external systems, the periodic R01 message sent by Gateway at the time does not have
NIBP data.