18
Repeat
Keep
8.0 FLUE GAS ANALYSIS
8.1 FLUE GAS ANALYSIS
To perform complete flue gas analysis, follow the instructions below.
8.1.1 Switching on the instrument and auto-calibration
Keep
pressed for
a few
seconds
When the autozero phase is complete, push the key related to the interactive function “, to proceed
with the combustion analysis or push the key related to the interactive function “, to repeat the autozero
phase.
11/12/17
10:00
HARDWARE
Memories KO
Calibration KO
voltages
ADC channels
WARNING!
When the instrument is turned on, an hardware check
is performed on the memories and on the instrument
calibration. If some error occurs, they will be shown
with the activation of the Hardware
screen. In this
case, it is advisable to restart the instrument
. If the
problem keeps on appearing or frequently occurring
please contact the Seitron Service Center,
communicating the shown error.
E500
S.n.:0
Fw:1.00
SOME IMPORTANT WARNINGS TO CONSIDER DURING THE COMBUSTION ANALYSIS
ARE LISTED BELOW:
FOR A CORRECT ANALYSIS NO AIR MUST FLOW INTO THE PIPE FROM OUTSIDE DUE
TO A BAD TIGHTENING OF THE POSITIONING CONE OR A LEAK IN THE PIPELINE.
THE GAS PIPE MUST BE CHECKED IN ORDER TO AVOID ANY LEAKAGES OR OBSTRUCTIONS ALONG
THE PATH. THE CONNECTORS OF THE
GAS SAMPLING PROBE AND OF THE CONDENSATION FILTER
MUST BE WELL CONNECTED TO THE INSTRUMENT. KEEP THE CONDENSATION TRAP ON VERTICAL
POSITION DURING THE ANALYSIS; A WRONG POSITIONING MAY CAUSE CONDENSATE
INFILTRATIONS IN THE INSTRUMENT AND THUS DAMAGE THE SE
NSORS. AFTER EACH ANALYSIS
CHECK FOR ANY PRESENCE OF WATER IN THE CONDENSATE COLLECTION BOWL AND REMOVE IT
IF ANY. PUT THE PROBE BACK IN THE CASE ONLY AFTER YOU HAVE ELIMINATED CONDENSATE
FROM THE TUBE AND THE EXPANSION TANK (SEE CHAPTER 'MAINTENANCE
'). REPLACE THE FINE
DUST FILTER IF IT IS VISIBLY DIRTY OR WET (SEE CHAPTER 'MAINTENANCE'). DO NOT PERFORM ANY
MEASUREMENT WHEN THE FILTER IS REMOVED OR DIRTY IN ORDER TO AVOID ANY RISK OF
IRREVERSIBLE DAMAGES TO SENSORS AND ANALYZER ITSELF.
BEFORE TURNING ON THE INSTRUMENT:
CONNECT THE GAS SAMPLING PROBE TO THE INSTRUMENT.
STORING OF THE AMBIENT TEMPERATURE:
CONDENSING BOILERS / FURNACES: UPON COMPLETION OF THE AUTOZERO IN FRESH
OUTDOOR AIR, PRESS THE “KEEP” BUTTON TO STORE THE OUTDOOR TEMPERATURE
BEING USED FOR PRIMARY AIR.
ATMOSPHERIC BOILERS / FURNACES: UPON COMPLETION OF THE AUTOZERO IN FRESH
OUTDOOR AIR, MOVE THE E500 INDOORS TO THE EXACT LOCATION OF THE APPLIANCE
AND
ONCE THE TEMPERATURE STABILIZES TO THE INDOOR TEMPERATURE, PRESS THE
“KEEP”
BUTTON TO STORE THE INDOOR TEMPERATURE BEING USED FOR PRIMARY AIR.
INSERT THE TCK CONNECTOR OF THE
GAS SAMPLING PROBE BEFORE STARTING THE
INSTRUMENT AND LET IT DRAW CLEAN AIR, IN ORDER TO GET A CORRECT
TEMPERATURE
VALUE. IF THE CONNECTOR IS NOT PLUGGED IN, THE TEMPERATURE WILL NOT BE
ACQUIRED.
15/06/18 10:00
AUTOZERO
60
AutoZero in
Fresh
Air
15/06/18 10:00
AUTOZERO
AutoZero in Fresh
Air
T: 21.5°C
Keep
Repeat